1
Keysight ZA0060A Quick Start Guide
44
Test Steps and Description
Sensitivity Search
Performs PER test using the one of the three search modes below to sweep through the power values
set. Use this step to determine the lowest output power the DUT can support to meet the PER limit.
Step Settings
– Search Mode: Select one of the three options available.
– Binary Sensitivity (Fastest): Perform binary searches for the PER sensitivity level.
– Linear Sensitivity (Slowest): Search for the PER sensitivity level by sweeping from maximum to
minimum power. This search will stop when it discovers the ‘Last Pass’, the last PER value that
produces a ‘Pass’ verdict.
– Mid-point Bi-directional: Search for the PER sensitivity level using a combination of binary and
linear sensitivity searches. From the first break point of the binary search, it will sweep in a linear
pattern for the ‘Fail’ point and report the ‘Last Pass’ value.
The Early Fail Count, Delay between Step, Reconnect Retries, and Preferred Mode parameters are as
described under
test step.
(Midpoint of -40 and -80)
-40
-60
-80
-40
-50
-60
-50
-55
-60
Always add a midpoint in
between the two pass and fail
points.
-40
-60
-80
-40
-59 (stops here)
-60
Add a midpoint in the beginning
then start to sweep at 1dB step
from the pass to fail point.
Binary Search
Mid Point Sensitivity Search
Linear Sensitivity Search
High
Last Pass (stops here)
Low
(Midpoint)
It will stop when it reaches the
last value with a ‘Pass’.