1
Keysight ZA0060A Quick Start Guide
16
ii 34999-60009
USB cable from
2 (Micro-USB Port)
of 34999B to PC.
iii 34999-60011
USB cable from
3 (USB Port)
to
8 (USB Interface
Connector)
of the 34999B RF module.
3
Refer to
for instructions to set up the multi-up configuration.
4
Plug in the power cord to
7 (Power Line Fuse Holder Assembly)
and then to
the source of electricity.
5
Download and install the software listed in the next section, then power on the
34972A. Allow 15 seconds after powering on the 34972A for your PC to detect
the instrument and its VISA address.
NOTE
The input power measurement range of the ZA0060A low power WLAN option is
from -4 to -44 dBm at the RFIO port. WLAN devices can transmit at higher power
up to +36 dBm. Even after considering the path loss of cable/shield box, the
input power to the tester RFIO port may still be above the tester power range. An
external attenuator may be required to connect to the RFIO port to reduce input
power to within the tester power range
Keysight offers a variety of fixed attenuators (such as the 8493A/B/C) to choose
based on the output power of your device.