U3022AM42
Control Lines
Setting the Control Lines with Address and Data Values
This section describes how to control the rear panel DUT control lines.
The following control feature will function only while the analyzer is in
its Standalone mode.
Setting the Analyzer to Standalone Mode
1. Select
Utility
>
System
>
Configure
>
Multiport Capability
-OR-
(Instrument > Setup >
External
Hardware > Multiport > Multiport Configuration...)
.
In the dialog box select
Restart as a standalone PNA > OK.
2. Access Interface Control Mode (see
As described in
“Setting the Individual Test Port Paths with Address and Data” on page 45
, the
<address>.<data> values are determined in a similar manner, with the following exceptions:
•
Table 10 will be used.
•
The <address> value is always = 112
Table 1
1
Test Set DUT Control Address and Data Logic Table
<Address>
<Data>
Description
Line/Pin
112
0
ALL DUT Control Lines set to logic high or connected to Pin 13
1-8
112
255
ALL DUT Control Lines set to logic low or connected to Pin 14
1-8
112
1
DUT Control Line 1 set to logic low or connected to Pin 14
1
112
2
DUT Control Line 2 set to logic low or connected to Pin 14
2
112
4
DUT Control Line 3 set to logic low or connected to Pin 14
3
112
8
DUT Control Line 4 set to logic low or connected to Pin 14
4
112
16
DUT Control Line 5 set to logic low or connected to Pin 14
5
112
32
DUT Control Line 6 set to logic low or connected to Pin 14
6
112
64
DUT Control Line 7 set to logic low or connected to Pin 14
7
112
128
DUT Control Line 8 set to logic low or connected to Pin 14
8