U3022AM42
General Specifications
General Specifications
Specifications for the test set are characteristic for the system performance of the network
analyzer and test set. Actual performance of the system is based on the customer’s network
analyzer and options that are used with the test set. A functional certificate is only offered for
the test set.
An N-Port calibration should be performed for optimum measurement accuracy.
A periodic calibration is not required. The Operators Check should be performed after System
Setup, or if performance is in question.
When connected to a analyzer, the test set will degrade the performance at the test ports. The
internal switch paths reduce test port power to the receivers. This affects the test port power of
the analyzer and also reduces dynamic range.
CAUTION
This product is designed for use in Installation Category II and Pollution Degree 2.
Power Requirements
Verify that the required AC power is available before installing the test set to the analyzer.
• 100/120/220/240 VAC (50/60Hz)
• The instruments can operate with mains supply voltage fluctuations up to ± 10% of
the nominal voltage.
• Air conditioning equipment (or other motor–operated equipment) should not be placed on
the same AC line that powers the test set and analyzer.
• U3022AM42 maximum power is 350 W.
This is a Safety Class I product (provided with a protective earthing ground
incorporated in the power cord). The mains plug shall be inserted only into a
socket outlet provided with a protective earth contact. Any interruption of
the protective conductor inside or outside the instrument is likely to make
the instrument dangerous. Intentional interruption is prohibited.
Use Keysight supplied power cord or one with the same or better electrical
rating.
Keysight U3022AM42 User's and Service Guide