U3022AM42
Control Lines
Control Lines
The 15 pin female D-Sub connector on the rear panel provides 8 latched data lines that can be
used to control your device under test (DUT). The lines can be controlled with the Multiport
External Test Set control, or Test Set I/O commands. Refer to
and
“Setting the Control Lines with Address and Data Values” on page 54
See Table 8 below for DUT control line specification.
The output voltage of the lines can be from the internal adjustable voltage source (+2 to
+5 Vdc), or an external DC power supply depending on how the connection to the control line is
configured. When using an external power supply a positive or negative voltage can be used.
Refer to
and
for control line pin location and
description. Refer to
“Internal Voltage Supply Configuration” on page 52
and
Supply Configuration” on page 53
Table
9
DUT Control Specifications
Item
Specifications
Connector Shape
15–pin female D–Sub
Voltage Range:
Positive Input
0 to +5 V
Negative Input
–5 to 0 V
Maximum Current
100 mA (in total of each
line)
Impedance
< 10
Ω
Range of Variable Voltage
+2 to +5 V
50
Keysight U3022AM42 User's and Service Guide