System Operational Check
The
following is a raw measurement procedure to confirm that the RF interface cables between
the test set and analyzer are installed and the system is working correctly. Ensure that your
analyzer is operational before connecting the test set.
•
Stack the analyzer on top of the test set.
•
Install the Test Set I/O cable on the rear panel.
•
Install all of the RF Interface cables on the front panel.
•
Install the included female
3.5 mm
s
hort (
85052-60007
)
to Port 1 on the analyzer.
Equipment Setup
1.
Turn on the test set and the analyzer
.
2.
Preset the analyzer and make the following settings:
•
Set to Multi-Port Mode.
•
Frequency range: 10 MHz to 26.5 GHz
.
•
IFBW: 1 kHz, number of points: 201
•
Scale: 10 dB/Div
•
Set the PNA to measure S11
•
Ensure the RF path of the analyzer is in Default mode. Refer to
for Analyzer with Noise Figure
U3022AM42
System Operational Check
Complete the following steps before beginning this procedure:
Verify Results
T
he
included accessory
3.5 mm
female
short will be relocated to each Test
Port and an uncorrected reflection response trace should be displayed that
resembles the following plots indicated in Table
7
below. If you suspect an
RF signal path problem, refer to the RF cable diagram and check the
interface cables associated with this signal path for proper installation.
Table 7
Reflection Response Results
Reflection Port
Response Trace
Cable Path Diagram
Port 1
Port 2
Ports 3 to 14
Ports 15 to 18
The trace ripple (peak-peak variation) is normal for this raw uncorrected
measurement.