U3022AM42
Cal Kit Operational Check
Cal Set Verification
1. Select
Trace/Chan
>
Trace
>
Delete Trace
-OR-
(Instrument > Trace > Delete Trace)
.
There should be no traces on the screen.
2. To launch the Cal Set Viewer toolbar. Select
Response > Cal > Manage Cals > Cal Set Viewer
-OR-
(Response > Cal > Cal Sets & Cal Kits > Cal Set Viewer [On/Off])
.
Figure
4
6
Calibration Cal Set Viewer
3.
From the Cal Sets drop-down menu, select
999.1
and select
Enable
. Select the
Reflection
Tracking(x,x)
, where x,x is the port being tested. Ensure that the
Enable
and
Error
Terms
check boxes are selected.
4.
Compare the Reflection Tracking (1,1) trace to the appropriate limits in
Use the Markers function to measure the trace values.
5.
Repeat step 3 and step 4 for Cal Sets “999.1” thru “999.18” (18-Port).
60
Keysight U3022AM42 User's and Service Guide