U3022AM42
Controlling the Test Set with N5242A/B
Trace Measure S-Parameter
S-Parameter selection can be accomplished using
Response > Measure
. Use the drop-down
menu to select 1 of 400 S-Parameters for the 20-Port system. The first number in the Sxx
selection is the Receiver Port and the second number will be the Source Port. Any port can be
selected to be the Receiver, Source or both, as in S11. The front panel R LED indicates the port
is the Receiver and the S LED indicates the port is the Source.
The 'B' model analyzer provides an S-Parameter selection method by using the keypad. For
example, to select S25:
MEAS [MEAS] > Enter S-Parameters > [2][.][5] [Enter]
Figure
2
6
S-Parameter Measurement ('A' Model Analyzers Only)
New Trace Measure S-Parameter
S-Parameter Tab:
Multiple S-Parameters can be made from the New Trace
menu. In the drop-down
menu select
Trace/Chan > New Trace
-OR-
(Instrument > Trace > New Trace)
.
The dialog box allows the selection of any of the 400 S-Parameters.
Figure
2
7
New S-Parameter Measurement
Keysight U3022AM42 User's and Service Guide
33