The Keysight U3022AH10 High Power Test Set can be configured for many applications.
Included in this document are four typical configurations:
•
•
“1 Watt High Power Configuration”
•
, “20 Watt High Power Configuration”
The internal firmware of the
Keysight
N5242A
/B
PNA-X Series Network
Analyzer has not been modified for this Test Set option. Power levels may
differ from those indicated on the PNA-X when the Test Set is connected.
Refer to the configuration diagrams for external component connections and/or
operating constraints when utilizing the high power capability. External
components are not supplied with the high power Test Set
,
found on
The high power Test Set is equipped with receiver access ports. The jumpers can
be removed to insert attenuators that will reduce the RF power to the PNA-X A,B
and R channel receivers. Recommended power levels to the PNA-X receiver
ports is –15 dBm. Refer to your PNA-X specifications to optimize power levels to
the receiver ports.
Prior to powering-up the booster amplifier it is highly recommended that the
user verify the RF power levels seen by the various elements of the Test Setup.
Both the PNA-X and Test Set power levels must be taken into consideration. At
high power levels, a mistake could damage the instruments. Accordingly, the
following key values are given in
U3022AH10 User's and Service Guide