Setting Option 129 Source Path
1.
Restart the PNA to Standalone Mode, if it is in Multiport Mode. Refer to
2.
Send command 0.0 and [Preset] the PNA-X.
3.
"Interface Control Mode" on page
set Port 1, 3, 5, 7, 9, or 11 as the noise source path.
The test set will not operate in multiport mode. Interface or
SCPI control
mode
is required to set the Noise Source Path.
Table
10
Option 129 Source Path Test Set I/O Commands
Address Data
Source Path
0.0
Standard Port 1
(Option 129 is not enabled)
0.8
Noise Source Port 1
0.9
Noise Source Port 3
0.10
Noise Source Port 5
U3022AH10
Controlling the System with N5242A/B
0.11
Noise Source Port 7
0.12
Noise Source Port 9
0.13
Noise Source Port 11
An 8 is added to the Data Value to activate the Option 129 switch, as