/B
N-Port Calibration with N5242A
/B
It is recommended that you perform an ECal characterization to minimize the connections
required for multiple port calibration. Keysight cable (85131F) is recommended.
1.
On the network analyzer select Response > Calibration Wizard
-OR-
(Response > Cal > Other Cals > ECal...)
and follow the prompts.
a.
If using a mechanical cal kit, select SmartCal.
b.
If using an ECal module, select ECal.
2.
Continue following the Cal Wizard prompts. On the “Select Calibration Ports and ECal
Module” window, press Select All or select the ports you are calibrating and press Next.
Figure 33 1-Port Calibration Selection
3. Connect the ECal or the mechanical cal kit to the ports you are calibrating following the Cal
Wizard prompts and press Measure after each connection.
4. At the Calibration Completed prompt, select Save As User Calset and type the name desired.
Press Save.
Help
40
U3022AH10 User's and Service Guide