General Specifications
U3022AH10
Frequency Range and Maximum Power Levels
The U3022AH10 frequency range is 10 MHz to 26.5 GHz.
It is recommend that you do not operate near damage levels. The power levels
must be 3 dB below maximum level to ensure no damage. See
The examples shown assumes a 10 dBm (0.01w) RF signal and 7 volt DC.
Equation 1 DC Wattage (
E
2
/R = DC Wattage)
Equation 2 Maximum Wattage (RF + DC = Maximum Wattage)
Table
6
Maximum Power Level Inputs
Test Set or PNA-X Front Panel
Connections
Power Level
1 Watt Standard
Configuration
1
Power Level
20 Watt
Configuration
2
Maximum U3022AH10 Multiport Test Set RF Power Levels:
PORT 3-12
+30 dBm 40 Vdc
+43 dBm 40 Vdc
SOURCE OUT, Noise Source In/Out
+30 dBm 0 Vdc
+30 dBm 0 Vdc
CPLR ARM
+30 dbm 7 Vdc
+30 dbm 7 Vdc
CPLR THRU
+30 dBm 40 Vdc
+43 dBm 40 Vdc
RVCR A IN and B IN
3
+15 dBm 0 Vdc
+15 dBm 0 Vdc
RVCR R1 or R2 IN
3
+15 dBm 7 Vdc
+15 dBm 7 Vdc
Damage Power Levels to U3022AH10 Access and Test Ports:
4
1.
All jumpers installed are considered standard configuration.
2.
Isolators and attenuators installed. Refer to
3.
Refer to your PNA-X specifications to determine the maximum input power levels for the access
and test ports.
4.
Add the RF signal to the DC component to determine the maximum wattage.
Max Level to Port 1 & 2 Test Ports
+30 dBm 40 Vdc
+43 dBm 40 Vdc
0.01w + 0.98w = 0.99w
7v
2
/ 50Ω = 0.98w
U3022AH10 User's and Service Guide