:SECurity
:IMMediate
Clears all user memory and reboots the instrument.
:SET <block data>
Get and set the instrument state
:TIME <
hh
>, <
mm
>, <
ss
>
Sets the time of the system clock.
:VERSion?
Returns the SCPI version that the instrument complies with.
:ACQuire
[:IMMediate]
Generates an immediate trigger.
:CURRent
[:LEVel] <
value
>
Sets the triggered level of the output.
:SLOPe POSitive|NEGative Sets the slope of the signal.
:INDices
[:DATA]?
Returns the indices where triggers were captured.
:COUNt?
Returns the number of triggers captured during the acquisition.
:SOURce <
source
>
Selects the trigger source for the acquisition system: BUS |CURRent1 |EXTernal |PIN<1-7>
|TRANsient1 |VOLTage1
:TOUTput
[:ENABle] 0|OFF|1|ON
Enables measurement triggers to be sent to a digital port pin.
:VOLTage
[:LEVel] <
value
>
Sets the triggered level of the output.
:SLOPe POSitive|NEGative Sets the slope of the signal.
:ARB
:SOURce <
source
>
Selects the trigger source for arbitrary waveforms: BUS |EXTernal |IMMediate |PIN<1-7>
:ELOG
[:IMMediate]
Generates an immediate trigger.
:SOURce <
source
>
Selects the trigger source for external data logging: BUS |EXTernal |IMMediate |PIN<1-7>
:TRANsient
[:IMMediate]
Generates an immediate trigger.
:SOURce <
source
>
Selects the trigger source for the transient system: BUS |EXTernal |IMMediate |PIN<1-7>