Front Panel Menu Reference
SCPI Command
Select
Protect\UVP
Enter a value in the UVP level box.
Then check
Enable UVP
Enter a value in the UVP Delay box.
Then press
Select
.
To set the UVP level to 2 volts:
To enable low-voltage protection:
To specify a protection delay of 5 seconds:
Clear Output Protection
If an over-voltage, over-current, over-temperature, power-fail condition, power-limit condition,
protection condition, or inhibit signal occurs, the output is disabled. The appropriate operating status
indicator on the front panel will be on. To clear the protection function and restore normal operation,
first remove that condition that caused the protection fault. Then, clear the protection function as
follows:
Front Panel Menu Reference
SCPI Command
Select
Protect\Clear
Select Clear.
To clear a protection fault:
The output is re-enabled once the ouptut protection function is cleared.
Keysight RP7900 Series Operating and Service Guide
123
4 Using the Regenerative Power System