Front Panel Menu Reference
SCPI Command
Select
Transient\Arb\Config
.
If Arb points have been imported or programmed
using the SCPI command, the Points field displays
the number of points in the Arb.
View the amplitude of any Arb point by entering the
point number in the Point # field. The Level field
displays the amplitude.
To program 10 points in a current Arb:
ARB:CURR:CDW 1,2,2,3,4,4,3,2,2,1
To query the number of Arb points:
To query the Arb point values:
Specify how many times you want the Arb to repeat
Depending on your application, specify how many times you want the Arb to repeat. Sending the
INFinity parameter in the SCPI command makes the Arb repeat indefinitely. At reset, the Arb count is
set to 1.
Front Panel Menu Reference
SCPI Command
Select
Transient\Arb\Repeat
.
Enter the number of list repetitions (2) and press
Select.
To program the Arb to repeat twice:
Specify how you want the Arb to end
Specify the output state after the Arb has completed. There are two choices: the output returns to the
state it was in before the Arb started, or the output remains at the values of the last Arb point.
Front Panel Menu Reference
SCPI Command
Select
Transient\Arb\Terminate
.
Select either Return to Start, or Stop at Last Step
and press Select.
To return the output to the pre-Arb state:
To keep the output at the Arb end point:
Keysight RP7900 Series Operating and Service Guide
133
4 Using the Regenerative Power System