VOLTage Subsystem
Voltage commands program the output voltage of the instrument.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <
value
>|MIN|MAX
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX]
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <
value
>|MIN|MAX
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX]
Sets the immediate or triggered voltage level when the output is operating in voltage priority mode.
The triggered level is the value that is transferred to the output when an output Step is triggered.
Units are in volts. The maximum value depends on the voltage rating of the unit.
Parameter
Typical Return
0.1% to 102% of rating, *RST 0.1% of rating
<voltage level>
Sets the output voltage to 20 V: VOLT 20
Sets the triggered voltage to 10 V: VOLT:TRIG 10
[SOURce:]VOLTage:BWIDth LOW|HIGH1
[SOURce:]VOLTage:BWIDth?
This command is provided for backward compatibility with previous RP795xA and
RP796xA units. Newer units should use the VOLT:BWID:RANG command.
Sets the voltage compensation. This lets you optimize output response time with capacitive loads.
These compensation modes only apply when the unit is operating in voltage priority mode.
LOW (0) is optimized for stability with a wide range of output capacitors.
HIGH1 (1) provides maximum up-programming speed as well as the fastest transient response time
when the output capacitance is restricted to small values. Refer to
for
specific capacitive load limits.
Parameter
Typical Return
LOW|HIGH1, *RST HIGH1
LOW or HIGH1
Sets the voltage bandwidth to Low: VOLT:BWID LOW
[SOURce:]VOLTage:BWIDth:LEVel 0 | 1 | 2, <
value
>|MIN|MAX
[SOURce:]VOLTage:BWIDth:LEVel? 0 | 1 | 2, [MIN|MAX]
Sets the programming pole frequency associated with each voltage priority compensation range. The
value is in Hertz. The default frequency is optimized for maximum up-programming speed as well as
the fastest transient response time. It can be reduced to compensate for output overshoots.
For RP795xA and RP796xA units, you can only specify two compensation ranges 0 (Low) or 1 (High1).