*TRG
Trigger command. Generates a trigger when the trigger subsystem has BUS selected as its source.
The command has the same effect as the Group Execute Trigger (<GET>) command.
Parameter
Typical Return
(none)
(none)
Generates an immediate trigger: *TRG
*TST?
Self-test query. Performs a instrument self-test. If self-test fails, one or more error messages will
provide additional information. Use SYSTem:ERRor? to read error queue. See
for more information.
Parameter
Typical Return
(none)
0 (pass) or
+1 (failed)
Perform self-test: *TST?
l
The power-on self-test is the same self-test performed by *TST.
l
*TST? also forces an *RST command.
*WAI
Pauses additional command processing until all pending operations are complete. See
for more
information.
Parameter
Typical Return
(none)
(none)
Wait until all pending operations complete. *WAI
l
*WAI can only be aborted by sending the instrument a Device Clear command.
5 SCPI Programming Reference
200
Keysight RP7900 Series Operating and Service Guide