MSP
There has been a communication loss between the master and slave units. Check that the
master/slave
are installed correctly. They should not be installed in the LAN or SDS
connector.
CSF
There has been a current sharing fault due to the master/slave loop being out of regulation. This can
occur if the
are set incorrectly. It can also occur temporarily if there has been a
mode crossover from a load transient or due to rapid up/down programming. If this is the case, set the
current limit to a higher value. CSF can also occur if the loop can no longer regulate the current
imbalance between units. CSF can also be an initial indication of communication loss with the master
unit; which will transition to MSP.
If the following error messages occur:
No slaves discovered or 1 slave discovered (if expecting multiple slave units)
Check that there are no duplicate slave addresses.
Error 332, Master/Slave Error
This most often indicates that the master unit has lost communication with a slave, and is usually
accompanied by MSP. Check that the master/slave
are installed correctly. This can also
occur when changing the master slave configuration. Always set the master unit to None when
changing master/slave configurations.
Independent Parallel Operation - RP795xA, RP796xA
Never leave any units powered on while the others are off. If AC power is turned on or
off on a single unit, the outputs of the units that are on will be enabled and will
transition to their programmed settings. To prevent this, program the outputs of all
units to zero before powering them off, and always power all units on and off together.
Enable Current Sharing
The current sharing function is automatically enabled for models RP795xA, RP796xA
in master/slave operation.
Units can be operated in parallel without current sharing, but the output current will
not be shared equally, and constant voltage mode may not be maintained on all units.
Current sharing is an analog control function that fine-tunes the output voltage up to about 0.5% of
the unit's voltage rating. This improves the current readback accuracy among paralleled units. This
applies whether units are connected for master/slave operation or whether they are connected for
independent parallel operation.
Keysight RP7900 Series Operating and Service Guide
115
4 Using the Regenerative Power System