![Keysight Technologies N69100 Series Operating And Service Manual Download Page 238](http://html1.mh-extra.com/html/keysight-technologies/n69100-series/n69100-series_operating-and-service-manual_1962218238.webp)
Current Programming and Measurement Calibration
Step 1. Disconnect all equipment form the output terminals.
Step 2. Select the current programing and measurement calibration.
Front Panel Menu Reference
SCPI Command
Select
System\Admin\Cal\Curr\Iprog
.
Check that nothing is connected to the output and select
Next.
Specify the full-scale current range. Full scale
ranges vary by model. This selects the 30 A range:
Step 3. Wait 5 minutes for the temperature to settle. Then select the first current calibration point.
Front Panel Menu Reference
SCPI Command
Wait 5 minutes, then select Next again.
Step 4. Connect a precision shunt resistor to the output. The shunt resistor should be able to measure
at least 70% of the output’s full-scale current (see
). Connect the Keysight 3458A DMM
across the shunt resistor.
Front Panel Menu Reference
SCPI Command
Check that the shunt is connected and select Next.
Not applicable
Step 5. Select the second current calibration point. Wait 5 minutes for the temperature to settle.
Calculate the shunt current (I=V/R) and enter the data.
Front Panel Menu Reference
SCPI Command
Display shows: “Enter P2 measured data”. Enter the data from
the external DMM. This should be about 70% of the full-scale
current rating. Press Enter when done. Press Back to finish.
CAL:LEV P2
*OPC?
CAL:DATA <data>
Enter a Calibration Date
Front Panel Menu Reference
SCPI Command
Select
System\Admin\Cal\Date
.
Enter the calibration date in the Date field. If desired, you
can enter alphanumeric data in this field.
Save Calibration and Log Out
Front Panel Menu Reference
SCPI Command
Select
System\Admin\Cal\Save
.
Select Save to save all calibration data.
Select
System\Admin\Logout
to exit calibration mode.
To save calibration data:
To exit calibration mode:
6 Verification and Calibration
238
Keysight N69100 Series Operating and Service Guide