Capacitive Load Boundary for Constant Voltage (CV) Operation
The following figures show the boundary limitations for capacitive loads with equivalent series
resistance (ESR) for constant voltage operation. Operation below the marginal stability lines with the
bandwidth settings may result in output instability. There are no absolute limits for DUT
capacitance, as the stability is also a function of ESR. Refer to
Load Capacitance and Lead Inductance
for details.
Capacitive Load Boundary (N69172A units, Comp 0-2 )
Capacitive Load Boundary (N69173A units, Comp 0-2 )
1. The output is stable in this region
2. The output may overshoot in this region
3. The output may be unstable in this region