Cannot initiate transient generator. Either the voltage or current function is set to Fixed mode.
310 The command is not supported by this model
This instrument does not have either the hardware or options required to support this command.
315 Settings conflict error
A data element could not be programmed because of the present instrument state.
316 Mass storage error
The mass storage memory has been exceeded.
317 Invalid format
An invalid data format was found in the command string.
320 Firmware update error
This may be due to the instrument hardware not being able to support the firmware version.
324 Inconsistent arb settings
The arb settings are inconsistent; most likely a mismatch in the arb lengths.
327 Initiated with no sense function enabled
A measurement has been initiated without specifying the measurement (sense) function.
328 Too many measurement points
Too many measurement points have been specified.
331 Illegal parameter value
The parameter value is out of range or does not exist.
332 Primary/secondary error
An error has occurred in the primary/secondary configuration
Command Errors (these errors set Standard Event Status register bit #5)
-100 Command error
Generic syntax error.
-101 Invalid character
An invalid character was found in the command string.
-102 Syntax error
Invalid syntax was found in the command string. Check for blank spaces.
-103 Invalid separator
5 SCPI Programming Reference
220