![Keysight Technologies N69100 Series Operating And Service Manual Download Page 203](http://html1.mh-extra.com/html/keysight-technologies/n69100-series/n69100-series_operating-and-service-manual_1962218203.webp)
VOLTage Subsystem
Voltage commands program the output voltage of the instrument.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude] <
value
>|MIN|MAX
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]? [MIN|MAX]
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] <
value
>|MIN|MAX
[SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]? [MIN|MAX]
Sets the immediate or triggered voltage level when the output is operating in voltage priority mode.
The triggered level is the value that is transferred to the output when an output Step is triggered.
Units are in volts. The maximum value depends on the voltage rating of the unit.
Parameter
Typical Return
0.1% to 102% of rating, *RST 0.1% of rating
<voltage level>
Sets the output voltage to 20 V: VOLT 20
Sets the triggered voltage to 10 V: VOLT:TRIG 10
[SOURce:]VOLTage:BWIDth:RANGe 0 | 1 | 2
[SOURce:]VOLTage:BWIDth:RANGe?
Sets the voltage compensation range. This lets you optimize output response time with capacitive
loads. These compensation modes only apply when the unit is operating in constant voltage (CV),
both in voltage priority and in current priority (when in voltage limit) . Refer to
for information about the effect that the compensation settings have on the Programming
Speed Characteristics.
0 (High speed/Small capacitive load) – provides the fastest programming speed and transient
response time.
1 (Medium speed/Medium capacitive load) – provides intermediate programming speed and transient
response time. setting is optimized for stability with a wide range of output capacitors.
2 (Slow speed/Large capacitive load) – best suited for DUTs with high capacitance/low ESR with the
trade-off of a slower programming speed and transient response.
Parameter
Typical Return
0 | 1 | 2, *RST 0
0, 1, or 2
Sets the voltage bandwidth to comp 1: VOLT:BWID:RANG 1
[SOURce:]VOLTage:LIMit[:POSitive][:IMMediate][:AMPLitude] <
value
>|MIN|MAX
[SOURce:]VOLTage:LIMit[:POSitive][:IMMediate][:AMPLitude]? [MIN|MAX]
Sets the voltage limit when in current priority mode. Units are in volts.
Keysight N69100 Series Operating and Service Guide
203
5 SCPI Programming Reference