50
2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Test External Trigger Sensitivity
This test applies to all models.
Verify the external trigger sensitivity at these settings:
•
100 MHz, 200 mV
pp
•
200 MHz, 350 mV
pp
1
Connect the equipment (see
a
Use the N cable to connect the signal generator to the power splitter input.
b
Connect one output of the power splitter to the Aux Trig input through a 50
Ω
feedthrough termination.
c
Connect the power sensor to the other output of the power splitter.
Summary of Contents for InfiniiVision 2000
Page 1: ...Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide...
Page 8: ...8 2000 3000 X Series Oscilloscopes Service Guide Contents...
Page 10: ...10 2000 3000 X Series Oscilloscopes Service Guide Figures...
Page 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Tables...
Page 62: ...62 2000 3000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 72: ...72 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies...