2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Table 8
Settings Used to Verify DC Vertical Gain Accuracy, 2000 X-Series Models
Volts/Div Setting
Precision Source Setting
Test Limits
5 V/Div
35 V
33.8 V
to
36.2 V
2 V/Div
14 V
13.52 V
to
14.48 V
1 V/Div
7 V
6.76 V
to
7.24 V
500 mV/Div
3.5 V
3.38 V
to
3.62 V
200 mV/Div
1.4 V
1.352 V
to
1.448 V
100 mV/Div
700 mV
676 mV
to
724 mV
50 mV/Div
350 mV
338 mV
to
362 mV
20 mV/Div
140 mV
135.2 mV
to
144.8 mV
10 mV/Div
70 mV
67.6 mV
to
72.4 mV
5 mV/Div
1
35 mV
33.4 mV
to
36.6 mV
2 mV/Div
1, 2
14 mV
12.72 mV
to
15.28 mV
1 mV/Div
1, 2
7 mV
5.72 mV
to
8.28 mV
1
A blocking capacitor is required at this range to reduce noise. See
2
Full scale is defined as 32 mV on the 2 mV/div and the 1 mV/div range. Full scale on all other ranges
is defined as 8 divisions times the V/div setting.