Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
35
To verify dual cursor accuracy
This test verifies the dual cursor accuracy for each analog channel.
This test is similar to the test for verifying the DC vertical gain, except you will
measure the dc voltage output of a precision source using dual cursors on the
oscilloscope and compare the results with the multimeter reading.
Dual cursor accuracy test limits: ±[DC vertical gain ac 0.5% full scale]
For the DC vertical gain accuracy test limits, see
1
Press
[Save/Recall] > Default/Erase > Factory Default
to recall the factory default
setup.
2
If you are testing a 2000 X-Series oscilloscope, set the probe attenuation to 1:1
on the analog channel you are testing (for example,
[1] > Probe > Probe
; then,
turn the
Entry knob to select
1.00 : 1
).
3
Set up the oscilloscope.
a
Set the Volts/Div setting to the value in the first line in
(depending on the oscilloscope model).
b
Adjust the channel 1 position knob to place the baseline at 0.5 major division
from the bottom of the display.
Table 10
Equipment Required to Verify Dual Cursor Accuracy
Equipment
Critical Specifications
Recommended Model/Part
Precision Source
DC voltage of 7 mV to 35 V, 0.1 mV resolution
Keysight B2912A/B2962A
Digital multimeter
Better than 0.01% accuracy
Keysight 34401A/34461A
50
Ω
BNC Cable (qty 2)
BNC - BNC, 48” length
Keysight 8120-1840
Adapter (qty 2)
BNC (f) to banana (m)
Keysight 1251-2277
BNC Tee
BNC tee (m) (f) (f)
Keysight 1250-0781 or Pomona 3285
Shorting cap
BNC
Keysight 1250-0774
Blocking capacitor
Note: if a BNC blocking capacitor is not available use
an SMA blocking capacitor.
Keysight
Pomona 4289 +
Pomona 5088
Summary of Contents for InfiniiVision 2000
Page 1: ...Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide...
Page 8: ...8 2000 3000 X Series Oscilloscopes Service Guide Contents...
Page 10: ...10 2000 3000 X Series Oscilloscopes Service Guide Figures...
Page 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Tables...
Page 62: ...62 2000 3000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 72: ...72 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies...