Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
47
To verify trigger sensitivity
This test verifies the trigger sensitivity. In this test, you will apply a sine wave to
the oscilloscope at the upper bandwidth limit. You will then decrease the
amplitude of the signal to the specified levels, and check to see if the oscilloscope
is still triggered.
Test limits for:
• Internal trigger sensitivity on all models:
•
< 10 mV/div: greater of 1 div or 5 mV
pp
•
>= 10 mV/div: 0.6 div
• External trigger sensitivity on all models:
•
DC to 100 MHz: < 200 mV
pp
•
100 MHz - 200 MHz: < 350 mV
pp
Table 16
Equipment Required to Verify Trigger Sensitivity
Equipment
Critical Specifications
Recommended Model/Part
Signal Generator
25 MHz, 100 MHz, 350 MHz, 500 MHz, and 1 GHz sine
waves
Keysight N5171B
Power splitter
Outputs differ < 0.15 dB
Keysight 11667A
Power Meter
Keysight N1914A
Power Sensor
Keysight E9304A or N8482A
Cable
Type N (m) 24 inch
Keysight 11500B
Adapter
Type N (m) to BNC (m)
Keysight 1250-0082 or
Pomona 3288 with Pomona 3533
50 Ohm Feedthrough
Termination
50
Ω
BNC (f) to BNC (m)
Keysight 0960-0301
Summary of Contents for InfiniiVision 2000
Page 1: ...Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide...
Page 8: ...8 2000 3000 X Series Oscilloscopes Service Guide Contents...
Page 10: ...10 2000 3000 X Series Oscilloscopes Service Guide Figures...
Page 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Tables...
Page 62: ...62 2000 3000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 72: ...72 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies...