Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
41
1
Connect the equipment (see
a
Use the N cable to connect the signal generator to the input of the power
splitter input.
b
Connect the power sensor to one output of the power splitter.
c
Use an N-to-BNC adapter to connect the other splitter output to the
channel 1 input.
Figure 7
Setting Up Equipment for Bandwidth (-3 dB) Verification Test
Power Splitter
Power Sensor
Signal
Generator
Oscilloscope
Power
Meter
N Cable
N to BNC Adapter