Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
29
1
Press
[Save/Recall] > Default/Erase > Factory Default
to recall the factory default
setup.
2
If you are testing a 2000 X-Series oscilloscope, set the probe attenuation to 1:1
on the analog channel you are testing (for example,
[1] > Probe > Probe
; then,
turn the
Entry knob to select
1.00 : 1
).
3
Set up the oscilloscope.
a
Adjust the horizontal scale to
200.0 us/div
.
b
Set the Volts/Div setting to the value in the first line in
(depending on the oscilloscope model).
c
Adjust the channel’s vertical position knob to place the baseline (reference
level) at 0.5 major division from the bottom of the display.
Table 7
Equipment Required to Verify DC Vertical Gain Accuracy
Equipment
Critical Specifications
Recommended Model/Part
Precision Source
DC voltage of 7 mV to 35 V, 0.1 mV resolution
Keysight B2912A/B2962A
Digital multimeter
Better than 0.01% accuracy
Keysight 34401A/34461A
50
Ω
BNC Cable (qty 2)
BNC - BNC, 48” length
Keysight 8120-1840
Adapter (qty 2)
BNC (f) to banana (m)
Keysight 1251-2277
BNC Tee
BNC tee (m) (f) (f)
Keysight 1250-0781 or Pomona 3285
Shorting cap
BNC
Keysight 1250-0774
Blocking capacitor
Note: if a BNC blocking capacitor is not available use
an SMA blocking capacitor.
Keysight
Pomona 4289 +
Pomona 5088
Summary of Contents for InfiniiVision 2000
Page 1: ...Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide...
Page 8: ...8 2000 3000 X Series Oscilloscopes Service Guide Contents...
Page 10: ...10 2000 3000 X Series Oscilloscopes Service Guide Figures...
Page 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Tables...
Page 62: ...62 2000 3000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 72: ...72 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies...