Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
31
d
Press the
[Acquire]
key.
e
Then press the
Acq Mode
softkey and select
Averaging
.
f
Then press the
#Avgs
softkey and set it to 64.
Wait a few seconds for the measurement to settle.
4
Add a measurement for the average voltage:
a
Press the
[Meas]
key.
b
Press
Source
; then, turn the Entry knob (labeled
on the front panel) to
select the channel you are testing.
c
Press
Type:
; then, turn the Entry knob to select
Average - Full Screen
, and press
Add Measurement
.
Table 9
Settings Used to Verify DC Vertical Gain Accuracy, 3000 X-Series Models
Volts/Div Setting
Precision Source Setting
Test Limits
5 V/Div
35 V
34.2 V
to
35.8 V
2 V/Div
14 V
13.68 V
to
14.32 V
1 V/Div
7 V
6.84 V
to
7.16 V
500 mV/Div
3.5 V
3.42 V
to
3.58 V
200 mV/Div
1.4 V
1.368 V
to
1.432 V
100 mV/Div
700 mV
684 mV
to
716 mV
50 mV/Div
350 mV
342 mV
to
358 mV
20 mV/Div
140 mV
136.8 mV
to
143.2 mV
10 mV/Div
70 mV
68.4 mV
to
71.6 mV
5 mV/Div
1
35 mV
34.2 mV
to
35.8 mV
2 mV/Div
1, 2
14 mV
13.36 mV
to
14.64 mV
1 mV/Div
1, 2
7 mV
6.36 mV
to
7.64 mV
1
A blocking capacitor is required at this range to reduce noise. See
2
Full scale is defined as 32 mV on the 2 mV/div range and the 1 mV/div range. Full scale on all other
ranges is defined as 8 divisions times the V/div setting.
Summary of Contents for InfiniiVision 2000
Page 1: ...Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide...
Page 8: ...8 2000 3000 X Series Oscilloscopes Service Guide Contents...
Page 10: ...10 2000 3000 X Series Oscilloscopes Service Guide Figures...
Page 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Tables...
Page 62: ...62 2000 3000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 72: ...72 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies...