Characteristics and Specifications
1
2000/3000 X-Series Oscilloscopes Service Guide
15
Maximum Input Voltage
CAUTION
Maximum input voltage for analog inputs
135 Vrms
CAUTION
Do not exceed 5 Vrms in 50
Ω
mode (3000 X-Series oscilloscopes only). Input protection
is enabled in 50
Ω
mode, and the 50
Ω
load will disconnect if greater than 5 Vrms is
detected. However, the input could still be damaged, depending on the time constant of
the signal. The 50
Ω
input protection mode only functions when the oscilloscope is
powered on.
CAUTION
When measuring voltages over 30 V, use a 10:1 probe.
CAUTION
Maximum input voltage for logic channels:
±40 V peak