44
2000/3000 X-Series Oscilloscopes Service Guide
2
Testing Performance
Then after converting the values from the power meter to Vrms:
16
The result from step 14 should be b3.0 dB and -3.0 dB. Record the
result in the Performance Test Record (see
17
Move the power splitter from the channel 1 to the channel 2 input.
18
Turn off the current channel and turn on the next channel using the channel
keys.
19
Repeat steps 3 through 17 for the remaining channels, setting the parameters
of the channel being tested where appropriate.
response(dB) = 20 log
10
mV
211.2
mV/
210.4
mV
185.3
/
mV
161.6
= -1.16 dB