Testing Performance
2
2000/3000 X-Series Oscilloscopes Service Guide
55
Keysight 3000 X-Series Oscilloscopes Performance Test Record
Serial No. ______________________________________
Test by _____________________________
Test Interval ____________________________________
Work Order No. ______________________
Recommended Next Testing ________________________
Temperature ____________
Threshold Specification
Limits
Ch
D7-D0
Ch
D15-D8
Accuracy Test
5 V - 250 mV
4.750 V
________
________
(100 mV + 3% of
threshold setting)
5 V + 250 mV
5.250 V
________
________
-5 V - 250 mV
-5.250 V
________
________
-5 V + 250 mV
-4.750 V
________
________
0 V - 100 mV
-100 mV
________
________
0 V + 100 mV
100 mV
________
________
DC Vertical Gain Accuracy
Range
Power Supply Setting
Test Limits
Channel 1 Channel 2 Channel 3* Channel 4*
5 V/Div
35 V
34.2 V to 35.8 V
________
________
________
________
2 V/Div
14 V
13.68 V to 14.32 V
________
________
________
________
1 V/Div
7 V
6.84 V to 7.16 V
________
________
________
________
500 mV/Div
3.5 V
3.42 V to 3.58 V
________
________
________
________
200 mV/Div
1.4 V
1.368 V to 1.432 V
________
________
________
________
100 mV/Div
700 mV
684 mV to 716 mV
________
________
________
________
50 mV/Div
350 mV
342 mV to 358 mV
________
________
________
________
20 mV/Div
140 mV
136.8 mV to 143.2 mV
________
________
________
________
10 mV/Div
70 mV
68.4 mV to 71.6 mV
________
________
________
________
5 mV/Div
35 mV
34.2 mV to 35.8 mV
________
________
________
________
2 mV/Div
14 mV
13.36 mV to 14.64 mV
________
________
________
________
1 mV/Div
7 mV
6.36 mV to 7.64 mV
________
________
________
________
Continued on next page.
Summary of Contents for InfiniiVision 2000
Page 1: ...Keysight InfiniiVision 2000 3000 X Series Oscilloscopes Service Guide...
Page 8: ...8 2000 3000 X Series Oscilloscopes Service Guide Contents...
Page 10: ...10 2000 3000 X Series Oscilloscopes Service Guide Figures...
Page 12: ...12 2000 3000 X Series Oscilloscopes Service Guide Tables...
Page 62: ...62 2000 3000 X Series Oscilloscopes Service Guide 3 Calibrating and Adjusting...
Page 72: ...72 2000 3000 X Series Oscilloscopes Service Guide 5 Replacing Assemblies...