2600S-901-01 Rev. C / January 2008
Return to
16-19
Series 2600 System SourceMeter® Instruments Reference Manual
Section 16: Calibration
Figure 16-4
Connections for contact check 0
Ω
calibration
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
LINE RATING
100-240VAC
50, 60Hz
240VA MAX.
DIGITAL I/O
RS-232
TSP-Link
LINE FUSE
SLOWBLOW
3.15A, 250V
CAT I
S
S
LO LO
HI
HI
G
G
G
G
CHANNEL A
!
!
!
UL
C
US
LISTED
SourceMeter
4ZA4
CAT I
S
S
LO
LO
HI
HI
G
G
G
G
CHANNEL B
!
R
Model 2602/2612 Channel A
connections shown
Output
LO
Sense
LO
Output
HI
Sense
HI
b. Allow the readings to settle, then get the Series 2600 readings:
r0_hi, r0_lo = smua.contact.r()
c. Connect
a
50
Ω
resistor between the sense low and output low terminals, as shown in
. Also connect a second 50
Ω
resistor between the sense high and output
high terminals as shown in the figure.
d. Allow the readings to settle, then get the Series 2600 readings:
r50_hi, r50_lo = smua.contact.r()
e. Send the contact check low calibration command:
smua.contact.calibratelo(r0_lo, Z_actual,r50_lo,50_ohm_actual)
Where:
r0_lo
Series 2600 0
Ω
low measurement
Z_actual
actual zero value
r50_lo
Series 2600 50
Ω
low measurement
50_ohm_actual
actual
50
Ω
resistor value
Typical values:
smua.contact.calibratelo(r0_lo,0,r50_lo,50.15)
f.
(Where
r0_lo
and
r50_lo
are measurements taken in steps 2 and 4 above.)
g. Send the contact check high calibration command:
smua.contact.calibratehi(r0_hi, Z_actual,r50_hi,50_ohm_actual)
Where:
r0_hi
Series 2600 0
Ω
high measurement
Z_actual
actual zero value
r50_hi
Series 2600 50
Ω
high measurement
50_ohm_actual
actual 50
Ω
resistor value
Typical values:
smua.contact.calibratehi(r0_hi,0,r50_hi,50.15)
(Where
r0_hi
and
r50_hi
are measurements taken in steps 2 and 4 above.)