2600S-901-01 Rev. C / January 2008
Return to
12-75
Series 2600 System SourceMeter® Instruments Reference Manual
Section 12: Instrument Control Library
Remarks
• This attribute controls the SMU action taken when the output enable or interlock line is
asserted/de-asserted. The default setting is
smuX.OE_NONE
.
• This attribute is always set to
OE_OUTPUT_OFF
for the Model 2611/2612 and is read-
only.
•
When set to
smuX.OE_NONE
, the SMU will take no action when the output enable or
interlock line goes low (de-asserted).
•
When set to
smuX.OE_OUTPUT_OFF
and the output enable or interlock line is de-
asserted, the SMU will turn its output off as if the
smuX.source.output =
smuX.OUTPUT_OFF
command had been received.
•
The SMU will not automatically turn its output on when the output enable or interlock
line returns to the high state.
• If the output enable or interlock line is not asserted when this attribute is set to
smuX.OE_OUTPUT_OFF
and the output is on, the output will turn off immediately.
• Detection of the output enable or interlock line going low will not abort any running
scripts. This may cause execution errors.
Details
See “
Output Enable (Models 2601/2602)
” in
Also see
Example
Reconfigures the SMU to turn the output off if the output enable or interlock line goes low
(de-asserted):
smua.source.outputenableaction = smua.OE_OUTPUT_OFF