2600S-901-01 Rev. C / January 2008
Return to
13-19
Series 2600 System SourceMeter® Instruments Reference Manual
Section 13: Factory Scripts
Waveform
Pulse initiated here
bias
level
toff
start
Measurement (1/NPLC seconds)
step
stop
ton
step = (stop - start)/(points -1)
Last source value before pulse train was initiated
step
Return
Values
f:
A Boolean flag. This flag will be
true
when the pulse was
successfully configured,
false
when errors were encountered.
msg:
A string message. If the
f
flag is
false
,
msg
will contain an error
message. Otherwise,
msg
will contain a string indicating successful
configuration.
Output
Data
Pulsed voltage measurements are stored in the reading buffer specified by the
buffer
input argument.
Example
ConfigPulseIMeasureVSweepLin(smua, 0, 0.01, 0.05, 1, 1e-3,
0.1, 20, smua.nvbuffer2, 3)
Set up a pulsed sweep that will use SourceMeter channel A. The pulsed sweep
will start at 10mA, end at 50mA, and return to a 0mA bias level between pulses.
Each pulsed-step will be on for 1msec and then at the bias level for 100msec. The
voltage limit will be 1V during the entire pulsed-sweep. The pulse train will be
comprised of 20 pulsed-steps and the pulse train will be assigned a tag index of 4.
See Also
,
Table 13-12 (continued)
KIGeneral TSP test script: ConfigPulseIMeasureVSweepLin
TSP project name:
KIFactoryPulse
TSP test script name:
KIPulse
Firmware version:
1.2.0 and later
Function:
ConfigPulseIMeasureVSweepLin
Usage:
f, msg = ConfigPulseIMeasureVSweepLin
(smu,bias,start,stop,limit,ton,toff,points,buffer,tag[,sync_in]
[,sync_out][,sync_in_timeout][,sync_in_abort])