Series 2600 System SourceMeter® Instruments Reference Manual
Section 12: Instrument Control Library
reset function
This function is used to return all logical instruments to the default settings.
Function
Resets the logical instruments to the default settings.
Usage
reset()
Remarks
•
This function resets all logical instruments in the system. This function is equivalent
to iterating over all the logical instruments in the system and calling the reset method
of each.
• Default settings are listed in
.
Details
See “
” in
.
serial functions and attributes
The functions and attributes in this group are used to configure the RS-232 Interface:
reset
serial.baud
Attribute
Baud rate for the RS-232 port.
Usage
baud = serial.baud
-- Reads baud rate.
serial.baud = baud
-- Writes baud rate.
baud
Set to 300, 600, 1200, 2400, 4800, 9600,
19200 38400, 57600 or 115200.
Remarks
• A new baud rate setting takes effect when the command to change it is processed.
• The user should allow ample time for the command to be processed before attempting to
communicate with the instrument again. It is recommended that the baud rate be set from
the GPIB interface or from the front panel.
• The baud rate is stored in non-volatile memory. The
reset
function has no effect on the
baud rate.
Details
See “
Also see
,
Example
Sets the baud rate to 1200:
serial.baud = 1200