16-12
Return to
2600S-901-01 Rev. C / January 2008
Section 16: Calibration
Series 2600 System SourceMeter® Instruments Reference Manual
24) Typical values for the Model 2611/2612/2635/2636 2V range:
smua.source.calibratev(-2,-1e-10,-1e-4,-1.8,-1.805)
25) If this step is not on the CALA sense mode, send the measure calibration command
using the multimeter and Series 2600 readings, and range setting for the
parameters:
26) smua.measure.calibratev(-range,Z_rdg,DMM_Z_rdg,
FS_rdg,DMM_FS_rdg)
Where:
-range
negative of the present calibration range
Z_rdg
-zero 2600 measurement
DMM_Z_rdg
-zero DMM measurement
FS_rdg
-FS 2600 measurement
DMM_FS_rdg
-FS DMM measurement
Typical Model 2601/2602 1V range values:
smua.measure.calibratev(-1,-1e-4,-1e-6,-0.89,-0.896)
27) Typical Model 2611/2612/2635/2636 2V range values:
smua.measure.calibratev(-2,-1e-4,-1e-6,-1.81,-1.805)
d. Be sure to complete steps a through v for all six voltage steps in
before continuing with current calibration.
e. Select auto polarity mode:
smua.cal.polarity = smua.CAL_AUTO