Replace N with the number of the synchronization line: 1 to 3.
Function
Waits for a trigger
.
Usage
triggered = tsplink.trigger[N].wait(timeout)
timeout
Specifies the time-out value in seconds.
triggered
A customized variable that stores the value true if a
trigger is detected, or false if a trigger is not
detected during the time-out period.
Remarks
This function waits up to
timeout
value in seconds for an input trigger. If one or more
trigger events were detected since the last t
ime
tsplink.trigger[N].wait
or
tsplink.trigger[N].clear
was
called, this function will return immediately. After
waiting for a trigger with this function, the event detector is automatically rearmed and reset.
This functionality is true regardless of the number of events detected.
Details
See
"TSP-Link Synchronization lines"
Also see
2600S-901-01 Rev. C / January 2008
Return to
12-115
Series 2600 System SourceMeter® Instruments Reference Manual
Section 12: Instrument Control Library
tsplink.trigger[N].wait
tsplink.writebit
Function
Sets a TSP-Link synchronization line high or low.
Usage
tsplink.writebit(bit, data)
N
The synchronization line number (1 to 3).
data
Value to write to the bit; 0 (low) or 1 (high).
Remarks
• If the output line is write protected by the
attribute, the
command will be ignored.
• The
reset
function does not affect the present states of the digital I/O lines.
• Use the
tsplink.writebit
and
tsplink.writeport
commands to control the
output state of the synchronization line when the trigger mode is set to
tsplink.TRIG_BYPASS
.
Details
See
"TSP-Link Synchronization lines"
Also see
,
Example
Sets synchronization line 3 low (0):
tsplink.writebit(3, 0)