12-110
Return to
2600S-901-01 Rev. C / January 2008
Section 12: Instrument Control Library
Series 2600 System SourceMeter® Instruments Reference Manual
tsplink function and attributes
Use the function and attributes in this section to assign node numbers to the Series 2600
instruments and initialize the TSP-Link system.
Details
See “
Also see
Example
Waits up to 10 seconds for a trigger:
triggered = trigger.wait(10)
print(triggered)
Output:
false
The above output indicates that no trigger was detected during the 10 second timeout.
tsplink.group
Attribute
Stores the group number assigned to the node.
Usage
groupnumber = tsplink.group
tsplink.group = groupnumber
groupnumber
The TSP-Link group number.
Remarks
• The default group number is 0.
• Set the attribute value to 0 to remove the node from all groups.
• Each time the node powers off, the group number for that node changes to 0.
• Group numbers can range from 0 to 64.
tsplink.master
Attribute
Reads the node number assigned to the master node.
Usage
master = tsplink.master
master
A custom variable that stores the master node
number.
Remarks
A read-only attribute.
tsplink.node
Attribute
Use this attribute to set the node number. Choose a number from 1 to 64.
Usage
mynode = tsplink.node
Reads the node number.
tsplink.node = mynode
Writes the node number.
mynode
Sets the node number. Choose from 1 to 64.
Remarks
• This attribute sets the TSP-Link node number and saves the value in nonvolatile memory.
• Changes to the node number does not take effect until the next time
tsplink.reset
is
executed on any node in the system.
• Each node connected to the TSP-Link must be assigned a different node number.
• 16 nodes is the maximum number of nodes allowed
Details
See
System expansion.
Also see
Example
--Sets the TSP-Link node to number 2:
tsplink.node = 2