Section 13: Factory Scripts
Series 2600 System SourceMeter® Instruments Reference Manual
Table 13-17
KIGeneral TSP test script: QueryPulseConfig
TSP project name:
KIFactoryPulse
TSP test script name:
KIPulse
Firmware version:
1.2.0 and later
Function:
QueryPulseConfig
Usage:
tbl = QueryPulseConfig(tag)
Description
Once a pulse train has been configured and assigned to a tag, it is often desirable
to inspect the settings of this pre-configured pulse train.
QueryPulseConfig()
can be used for this purpose. This function will return a table containing the
various settings associated with the tag input parameter.
Parameters
tag: Numeric identifier of pulse train configuration being queried.
Return
Values
tostring():
A function that returns most elements in a string convenient for
printing.
tag:
Identifying tag for this pulse train.
smu:
The smu configured for pulsing.
func:
Pulse function:
smuX.OUTPUT_DCAMPS
or
smuX.OUTPUT_DCVOLTS
bias:
Pulse bias level.
level:
Pulse level for non sweeping pulses.
start:
Starting level for sweep pulses.
stop:
Ending level for sweep pulses.
limit:
Limit value.
ton:
On time in seconds.
toff:
Off time in seconds.
points:
The number of points in this pulse train.
buf:
Reference to buffer containing measurement data
sync_in:
The sync_in digio line, if used.
sync_out:
The sync_out digio line, if used.
sourcevalues:
A table containing the source value for each point in the pulse
train.
Output
Data
None