Calibration 2-19
• Send the source and sense commands using the multimeter reading as the parameter.
For example:
:CAL:PROT:SOUR 1.997
:CAL:PROT:SENS 1.997
• Send the SOUR:VOLT 0.0 command to set the source voltage to 0V.
• Note and record the multimeter reading.
• Send the :CAL:PROT:SOUR command using the multimeter reading as the com-
mand parameter. For example:
:CAL:PROT:SOUR -1.02E-3
Table 2-8
Voltage calibration initialization commands
Command
Description
*RST
Restore GPIB defaults.
:SOUR:FUNC VOLT
Activate voltage source.
:SENS:CURR:PROT 0.1
Current limit when voltage source is active.
:SENS:CURR:RANG 0.1
Make sure 1A range is not active.
:SOUR:VOLT:PROT MAX
Maximum allowable source voltage.
:SYST:RSEN OFF
Disable remote sensing.
:CAL:PROT:CODE 'KI002430' Unlock cal.
:OUTP:STAT ON
Turn source on.
Step 3: Current calibration
1.
Connect the Model 2430 to the digital multimeter (Figure 2-2), and select the multimeter
DC current function.
2.
Send the commands summarized in Table 2-10 in the order listed to initialize current
calibration.
Summary of Contents for SourceMeter 2430
Page 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Page 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Page 16: ......
Page 17: ...1 Performance Verification ...
Page 33: ...2 Calibration ...
Page 56: ...2 24 Calibration ...
Page 57: ...3 Routine Maintenance ...
Page 60: ...3 4 Routine Maintenance ...
Page 61: ...4 Troubleshooting ...
Page 75: ...5 Disassembly ...
Page 82: ...5 8 Disassembly ...
Page 83: ...6 Replaceable Parts ...
Page 104: ...6 22 Replaceable Parts ...
Page 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Page 116: ...A 12 Specifications ...
Page 117: ...B Command Reference ...
Page 128: ...B 12 Command Reference ...
Page 129: ...C Calibration Programs ...
Page 140: ......