Calibration 2-15
3.
Select SAVE, then press ENTER. The unit will display the following message:
SAVE CAL
Press ENTER to continue; EXIT to abort calibration sequence
4.
Press ENTER. The unit will prompt you for the calibration date:
CAL DATE: 12/15/1998
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
5.
Change the displayed date to today's date, then press the ENTER key. Press ENTER
again to confirm the date.
6.
The unit will then prompt for the calibration due date:
NEXT CAL: 12/15/1998
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
7.
Set the calibration due date to the desired value, then press ENTER. Press ENTER again
to confirm the date.
8.
Once the calibration dates are entered, calibration is complete, and the following mes-
sage will be displayed:
CALIBRATION COMPLETE
Press ENTER to confirm; EXIT to abort
9.
Press ENTER to save the calibration data (or press EXIT to abort without saving cali-
bration data.) The following message will be displayed:
CALIBRATION SUCCESS
Press ENTER or EXIT to continue.
10.
Press ENTER or EXIT to complete process.
Step 5: Lock out calibration
1.
From normal display, press MENU.
2.
Select CAL, then press ENTER. The Model 2430 will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
©
ß
SAVE LOCK CHANGE-PASSWORD
3.
Select LOCK, then press ENTER to lock out calibration.
Remote calibration
Use the following procedure to perform remote calibration by sending SCPI commands over
the IEEE-488 bus or RS-232 port. The remote commands and appropriate parameters are sepa-
rately summarized for each step.
Remote calibration commands
Table 2-5 summarizes remote calibration commands. For a more complete description of
these commands, refer to Appendix B.
Summary of Contents for SourceMeter 2430
Page 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Page 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Page 16: ......
Page 17: ...1 Performance Verification ...
Page 33: ...2 Calibration ...
Page 56: ...2 24 Calibration ...
Page 57: ...3 Routine Maintenance ...
Page 60: ...3 4 Routine Maintenance ...
Page 61: ...4 Troubleshooting ...
Page 75: ...5 Disassembly ...
Page 82: ...5 8 Disassembly ...
Page 83: ...6 Replaceable Parts ...
Page 104: ...6 22 Replaceable Parts ...
Page 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Page 116: ...A 12 Specifications ...
Page 117: ...B Command Reference ...
Page 128: ...B 12 Command Reference ...
Page 129: ...C Calibration Programs ...
Page 140: ......