Command Reference B-3
Miscellaneous commands
Miscellaneous commands are those commands that perform such functions as saving calibra-
tion constants, locking out calibration, and programming date parameters.
:CODE
(:CALibration:PROTected:CODE)
Purpose
To unlock calibration so you can perform the calibration procedures.
Format
:cal:prot:code '<password>'
Parameter
Up to a 8-character string including letters and numbers.
Description
The :CODE command sends the password and enables calibration when per-
forming these procedures via remote. The correct password must be sent to
the unit before sending any other calibration command. The default remote
password is KI002430.
Note
• The :CODE command should be sent only once before performing
calibration. Do not send :CODE before each calibration step.
• To change the code, first send the present code, then send the new code.
• The password parameter must be enclosed in single quotes.
• If you change the first two characters of the password to something other
than “KI”, you will not be able to unlock calibration from the front panel.
Example
:CAL:PROT:CODE 'KI002430'
Send default code of KI002430.
:COUNT?
(:CALibration:PROTected:COUNT?)
Purpose
To request the number of times the Model 2430 has been calibrated.
Format
:cal:prot:count?
Response
Number of times calibrated.
Description
The :COUNT? query may be used to determine the total number of times the
Model 2430 has been calibrated.
Example
:CAL:PROT:COUNT?
Request calibration count.
:LOCK
(:CALibration:PROTected:LOCK)
Purpose
To lock out calibration.
Format
:cal:prot:lock
Query
:cal:prot:lock?
Response
0 Calibration unlocked
1 Calibration locked
Summary of Contents for SourceMeter 2430
Page 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Page 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Page 16: ......
Page 17: ...1 Performance Verification ...
Page 33: ...2 Calibration ...
Page 56: ...2 24 Calibration ...
Page 57: ...3 Routine Maintenance ...
Page 60: ...3 4 Routine Maintenance ...
Page 61: ...4 Troubleshooting ...
Page 75: ...5 Disassembly ...
Page 82: ...5 8 Disassembly ...
Page 83: ...6 Replaceable Parts ...
Page 104: ...6 22 Replaceable Parts ...
Page 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Page 116: ...A 12 Specifications ...
Page 117: ...B Command Reference ...
Page 128: ...B 12 Command Reference ...
Page 129: ...C Calibration Programs ...
Page 140: ......