Specifications
A-5
C O M P O N E N T HANDLER INTERFACE TIME:
4, 6
Speed
NPLC/Trigger Origin
Measure Pass/Fail Test
Source Pass/Fail Test
Source-Measure Pass/Fail Test
7
Fast
0.01 / external
1.01 ms (1.08 ms)
0.5 ms (0.5 ms)
5.3 ms
(5.3 ms)
Medium
0.10 / external
2.5 ms (2.9 ms)
0.5 ms (0.5 ms)
6.7 ms
(7.1 ms)
Normal
1.00 / external
17.5 ms (20.9 ms)
0.5 ms (0.5 ms)
21.7 ms (25.0 ms)
1
Reading rates applicable for voltage or current measurements. Auto zero off, autorange off, filter off, display off, trigger delay = 0, source auto clear off, and
binary reading format.
2
Purely resistive load. 10 A range <65ms.
3
1000 point sweep was characterized with the source on a fixed range.
4
Pass/Fail test performed using one high limit and one low math limit.
5
Includes time to re-program source to a new level before making measurement.
6
Time from falling edge of STA RT OF TEST signal to falling edge of end of test signal.
7
Command processing time of :SOURce:VOLTage|CURRent:TRIGgered <nrf> command not included.
General
NOISE REJECTION:
N P L C
N M R R
C M R R
Fast
0.01
—
80 dB
Medium
0.1
—
80 dB
Normal
1
60 dB
120 dB
1
1
Except lowest current range = 90dB.
LOAD IMPEDANCE: Stable into 20,000pF typical.
C O M M O N M O D E V O LTAGE: 250VDC, externally limited to 3.15A max.
COMMON MODE ISOLATION: >10
9
ž, <1000pF.
OVERRANGE: 105% of range, source and measure.
MAX. VOLTAGE DROP BETWEEN INPUT/OUTPUT AND SENSE TERMINALS: 5 volts
.
SOURCE/MEASURE TERMINAL R ATING: Installation cetegory I.
MAX. SENSE LEAD RESISTANCE: 1Mž for rated accuracy.
SENSE INPUT IMPEDANCE: >10
10
ž
.
GUARD OFFSET V O LTAGE: 500 V, typical.
SOURCE OUTPUT MODES:
Pulse
Memory List (mixed function)
Fixed DC level
Stair (linear and log)
S O U R C E M E M O RY LIST:100 points max.
M E M O RY BUFFER: 5,000 readings @ 5H digits (two 2,500 point buffers). Includes selected measured value(s) and time stamp. Lithium battery
backup (3 yr+ battery life).
PROGRAMMABILITY:IEEE-488 (SCPI-1996.0), RS-232, 5 user definable power-up states plus factory default and *RST.
DIGITAL INTERFACE:
Safety Interlock: Active low input.
Handler Interface: Start of test, end of test, 3 category bits. +5V @ 300mA supply.
Digital I/O: 1 trigger input, 4 TTL/Relay Drive outputs (33V @ 500mA sink, diode clamped).
POWER SUPPLY:100V to 240V rms, 50–60Hz, 250VA.
COOLING: Forced air, variable speed.
W A R R A N T Y:1 year.
EMC: Conforms with European Union Directive 89/336/EEC EN 61326.
SAFETY:
Conforms with European Union Directive 73/23/EEC EN 61010-1.
VIBRATION: MIL-T-28800E Type III, Class 5.
W ARM-UP: 1 hour to rated accuracies.
DIMENSIONS: 89mm high
∞
213mm wide
∞
370mm deep (3
1
¼
2
in
∞
8
3
¼
8
in
∞
14
9
¼
16
in). Bench Configuration (with handle & feet): 104mm high
∞
238mm wide
∞
370mm deep (4
1
¼
8
in
∞
9
3
¼
8
in
∞
14
9
¼
16
in).
WEIGHT:4.1kg (9.0 lbs).
ENVIRONMENT:Operating: 0°–50°C, 70% R.H. up to 35°C. Derate 3% R.H./°C, 35°–50°C. Storage: –25°C to 65°C.
Specifications subject to change without notice.
Section A (pgs 1-5).qxd 11/7/2005 1:35 PM Page 5
Summary of Contents for SourceMeter 2430
Page 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Page 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Page 16: ......
Page 17: ...1 Performance Verification ...
Page 33: ...2 Calibration ...
Page 56: ...2 24 Calibration ...
Page 57: ...3 Routine Maintenance ...
Page 60: ...3 4 Routine Maintenance ...
Page 61: ...4 Troubleshooting ...
Page 75: ...5 Disassembly ...
Page 82: ...5 8 Disassembly ...
Page 83: ...6 Replaceable Parts ...
Page 104: ...6 22 Replaceable Parts ...
Page 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Page 116: ...A 12 Specifications ...
Page 117: ...B Command Reference ...
Page 128: ...B 12 Command Reference ...
Page 129: ...C Calibration Programs ...
Page 140: ......