Command Reference B-5
:NDUE
(:CALibration:PROTected:NDUE)
Purpose
To send the next calibration due date to the instrument.
Format
:cal:prot:ndue <year>, <month>, <day>
Parameters
<year> = 1994 to 9999
<month> = 1 to 12
<day> = 1 to 31
Query
:cal:prot:ndue?
Response
<year>, <month>, <day>
Description
The :NDUE command allows you to store the date when calibration is next
due in instrument memory. You can read back the next due date by using the
:NDUE? query, or by using the front panel CAL menu.
Note
The next due date parameters must be delimited by commas.
Example
:CAL:PROT:NDUE 1999,11,20
Send due date (11/20/99).
:SENSE
(:CALibration:PROTected:SENSe)
Purpose
To calibrate the sense function.
Format
:cal:prot:sens <nrf>
Parameters
See Table B-2 for recommended parameter ranges for optimum calibration.
Description
The :CAL:PROT:SENS command calibrates the Model 2430 sense function.
To use this command, you must (1) program the source to the correct value,
(2) select the range being calibrated, and (3) send the :CAL:PROT:SENS
command for each parameter listed in Table B-2.
Note
When the Model 2430 receives this command, it will attempt to place the pa-
rameter into one of three parameter ranges summarized in Table B-2, de-
pending on the active :SENS:FUNC and range. If the parameter does not fit
into any of the three allowed ranges, an error number -222, “Parameter data
out of range” will be generated. Once the unit has successfully selected the
appropriate parameter range, it will then check to see if autorange for the ac-
tive :SENS:FUNC is enabled. If so, an error number -221, “Settings conflict”
will be generated. If no error occurs, the active sense function/range point
will be calibrated using the corresponding parameter.
Note that parameters for a given sense function and range may be sent in any
order; however, once one :CAL:PROT:SENS command executes, the other
two must also be executed, or an error number -200, “Execution error” will
occur when the :CAL:PROT:SAVE command is processed at the end of the
calibration procedure.
Example
:CAL:PROT:SENS 2
Calibrate 2V sense range.
Summary of Contents for SourceMeter 2430
Page 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Page 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Page 16: ......
Page 17: ...1 Performance Verification ...
Page 33: ...2 Calibration ...
Page 56: ...2 24 Calibration ...
Page 57: ...3 Routine Maintenance ...
Page 60: ...3 4 Routine Maintenance ...
Page 61: ...4 Troubleshooting ...
Page 75: ...5 Disassembly ...
Page 82: ...5 8 Disassembly ...
Page 83: ...6 Replaceable Parts ...
Page 104: ...6 22 Replaceable Parts ...
Page 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Page 116: ...A 12 Specifications ...
Page 117: ...B Command Reference ...
Page 128: ...B 12 Command Reference ...
Page 129: ...C Calibration Programs ...
Page 140: ......