Model 2430
Digital Multimeter
Input LO
Amps
2430 1KW PULSE SourceMeter
250V
PEAK
5V
PEAK
HI
LO
OUTPUT
125V
PEAK
125V
PEAK
EDIT
TOGGLE
POWER
RANGE
INPUT/
OUTPUT
4- WIRE
SENSE
DISPLAY
ON/OFF
TERMINALS
FRONT/
REAR
AUTO
RANGE
!
®
EXIT
ENTER
CONFIG MENU
SWEEP
TRIG
REL
LOCAL
FILTER
LIMIT
DIGITS SPEED
V
Ω
MEAS
I
FCTN
V
I
SOURCE
2
3
0
1
6
7
8
9
4
+/-
5
STORE RECALL
EDIT
Figure 2-2
10
μ
A to 1A range
current calibration
test connections
Calibration 2-11
2.
From normal display, press the SOURCE I key.
3.
Press the EDIT key to select the source display field, and then use the down RANGE
key to select the 10
μ
A source range.
4.
From normal display, press MENU.
5.
Select CAL, and then press ENTER. The unit will display the following:
CALIBRATION
UNLOCK EXECUTE VIEW-DATES
©
ß
SAVE LOCK CHANGE-PASSWORD
6.
Select EXECUTE, then press ENTER. The instrument will display the following
message:
I-CAL
Press ENTER to 10.000
μ
A
7.
Press ENTER. The Model 2430 will 10
μ
A and simultaneously display the
following:
DMM RDG: +10.00000
μ
A
Use
ß
,
©
,
¹
,
ƒ
, ENTER or EXIT.
8.
Note and record the DMM reading, and then adjust the Model 2430 display to agree
exactly with the actual DMM reading. Use the up and down arrow keys to select the digit
value, and use the left and right arrow keys to choose the digit position (or use the num-
ber keys, 0-9, +/-). Note that the display adjustment range is within ±10% of the present
range.
9.
After adjusting the display to agree with the DMM reading, press ENTER. The instru-
ment will then display the following:
I-CAL
Press ENTER to 00.000
μ
A
Summary of Contents for SourceMeter 2430
Page 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031 ...
Page 6: ...Other brand names are trademarks or registered trademarks of their respective holders ...
Page 16: ......
Page 17: ...1 Performance Verification ...
Page 33: ...2 Calibration ...
Page 56: ...2 24 Calibration ...
Page 57: ...3 Routine Maintenance ...
Page 60: ...3 4 Routine Maintenance ...
Page 61: ...4 Troubleshooting ...
Page 75: ...5 Disassembly ...
Page 82: ...5 8 Disassembly ...
Page 83: ...6 Replaceable Parts ...
Page 104: ...6 22 Replaceable Parts ...
Page 105: ...A Specifications Section A pgs 1 5 qxd 11 7 2005 1 35 PM Page 1 ...
Page 116: ...A 12 Specifications ...
Page 117: ...B Command Reference ...
Page 128: ...B 12 Command Reference ...
Page 129: ...C Calibration Programs ...
Page 140: ......