17-6
Return to
6482-901-01 Rev. A / August 2012
Section 17: Calibration
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
15. Set the Model 5700A output to 19.0000
μ
A, and turn the output on.
16. Verify that the Model 6482 current reading is within the limits for the 20
μ
A range as shown in
.
17. Repeat steps 12 through 15 for the 200
μ
A through 20 mA ranges.
NOTE
The following steps verify the 20
μ
A through 20 mA ranges on channel 2.
18. With the power off, connect the Model 5700A to the Model 6482 channel 2 INPUT triaxial
connector as shown in
.
19. Repeat the procedure for channel 2 (step 13 through step 17). Make sure to select channel 2
by pressing the
MSR2
key.
Figure 17-1
Connections for current verification tests (2 nA to 2
μ
A ranges)
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
MADE IN
U.S.A.
INPUT/
OUTPUT
42V
PEAK
250V
PEAK
TRIGGER
LINK
4-WIRE
SENSE
HI
LO
LINE RATING
100-240VAC
50, 60, HZ
100VA MAX
RS232
IEEE-488
(ENTER IEEE ADDRESS
WITH FRONT PANEL MENU)
250V
PEAK
5V
PEAK
5V
PEAK
5V
PK
V, ,
GUARD
GUARD
SENSE
LINE FUSE
SLOWBLOW
2.5A, 250V
INTERLOCK-
DIGITAL I/O
REMOTE
PreAmp
!
250
PEAK
40V
PEAK
GUARD
IN/OUT
LOW
CA T I
MADE IN
U.S
.A
.
IN/OUT
HIGH
SENSE
MAINFRAME
250
PEAK
42V
PEAK
40V
PEAK
250
PEAK
WARNING:
NO INTERNAL OPERA TOR SER
VICEABLE
P AR
TS SERVICE BY Q
UALIFIED
PERSONNEL ONL Y
!
IN/OUT HIGH connector
Mainframe/preamp
cable
WARNING:
NO INTERNAL OPERATOR SERVICABLE PARTS,SERVICE BY QUALIFIED PERSONNEL ONLY.
CAUTION:
FOR CONTINUED PROTECTION AGAINST FIRE HAZARD,REPLACE FUSE WITH SAME TYPE AND RATING.
SERIAL NO.
120
LINE RATING
50, 60Hz
60 VA MAX
FUSE
LINE
630 mAT
(SB)
315 mAT
(SB)
100 VAC
120 VAC
220 VAC
240 VAC
CAT
I
MADE IN
U.S.A.
OUTPUT ENABLE
RS-232
TRIGGER LINK
INPUT
RATINGS MAX.
30 V @ 20mA
CHANNEL 1
CHANNEL 2
RATINGS MAX.
COMMON
MODE
200V
!
!
IEEE-488
(CHANGE IEEE ADDRESS
WITH FRONT PANEL MENU)
ANALOG
OUT
RATINGS MAX.
30 V @ 20mA
VOLTAGE
SOURCE
OUT
30 V @ 20mA
Model 6482
Model 6430
Channel 1
INPUT
Channel 2
INPUT
Model 4801
low noise
coaxial cable
Model 7078-TRX-BNC
3-slot male triax to
BNC adapter
Model 7078-TRX-GND
3-slot male triax to
BNC adapter (guard
removed)
Summary of Contents for 6482