16-20
Return to
6482-901-01 Rev. A / August 2012
Section 16: SCPI Command Reference
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
difference between the feed reading (see
CALC3 or CALC4 reading = feed reading - null offset
ACQuire
:CALCulate3:NULL:ACQuire
Acquire channel 1 REL value
:CALCulate4:NULL:ACQuire
Acquire channel 2 REL value
Description
These commands automatically acquire the null offset values for channel 1
(CALC3) and channel 2 (CALC4). The next available reading will become the
offset value.
Enable and read relative result
STATe <b>
:CALCulate3:NULL:STATe <b> Control channel 1 REL
:CALCulate4:NULL:STATe <b> Control channel 2 REL
Parameters
<b> =
1 or ON
Enable REL
0 or OFF
Disable REL
Query
:STATe?
Query state of REL
Description
These commands are used to enable or disable the null or offset value for REL.
When enabled, the CALC3 or CALC4 reading will include the null offset value
(see
). When disabled, the CALC3 or CALC4 reading will not include
the null offset.
DATA?
:CALCulate3:DATA?
Read channel 1 REL result
:CALCulate4:DATA?
Read channel 2 REL result
Description
These query commands are used to read the CALC3 or CALC4 REL result. When
CALC3 or CALC4 is enabled (see
), the returned value will include the
CALCulate5
The CALC5 commands configure and control the RATIO function. They perform the remote
equivalents to the front panel RATIO and CONFIG RATIO configuration menu.
Select RATIO calculation mode
FORMat <name>
:CALCulate5:FORMat <name> Select RATIO calculation mode
Parameters
<name> = C3C4
CALC3/CALC4
C4C3
CALC4/CALC3
Query
:FORMat?
Query selected RATIO calculation
Description
This command selects the RATIO calculation mode. With C3C4 selected, the
CALC3 result (channel 1 REL) is divided by the CALC4 result (channel 2 REL).
With C4C3 selected, the CALC4 result is divided by the CALC3 result.