6482-901-01 Rev. A / August 2012
Return to
3-11
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
Section 3: Basic Operation
Using the analog outputs
Each analog output provides a -10 V to +10 V DC output analogous to the input current. The
analog output signal will nominally source a ±10 V proportional, but inverted, signal to the input
current on all measurement ranges. For example, with a +1 mA current on the 2 mA measurement
range, the analog output voltage would be -5 V.
list typical analog output voltages for
various ranges and input currents.
Table 3-8
Analog output voltage examples
Range
Input current
Analog output
20 nA
+10 nA
-5 V
2 µA
-2 µA
+10 V
20 µA
+7 µA
-3.5 V
2 mA
+0.5 mA
-2.5 V
20 mA
-15 mA
+7.5 V