17-24
Return to
6482-901-01 Rev. A / August 2012
Section 17: Calibration
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
21. Select
I-CAL
and press the
ENTER
key. The instrument will display the following:
I-CAL CHANNEL SELECT
CHAN-1 CHAN-2
22. Select
CHAN-1
and press the
ENTER
key. The instrument will prompt for the positive full-range
current:
I (+FS)
1
= +20.00000 uA
Use
,
,
,
, ENTER or EXIT.
23. Set the Model 5700A current to +19.0000
μ
A.
24. Use the EDIT keys to adjust the Model 6482 display value to match the Model 5700A current.
25. Press the
ENTER
key. The instrument will prompt for the zero current:
I (0)
1
= +00.00000 uA
Use
,
,
,
, ENTER or EXIT.
26. Set the Model 5700A current to 0 nA and press the
ENTER
key. The instrument will prompt for
the negative full-range current:
I(-FS)
1
= -20.00000 uA
Use
,
,
,
, ENTER or EXIT.
27. Set the Model 5700A current to -19.0000
μ
A.
28. Adjust the displayed current to match the Model 5700A current, and then press the
ENTER
key
to complete calibration of the present range.
29. Press the
EXIT
key to return to normal display.
30. Repeat steps 18 through 29 for the 200
μ
A to 20 mA ranges, using
as a guide. Be
sure to set the Model 6482 to the correct range using the RANGE
and
keys, and use the
corresponding Model 5700A currents for each range.